A standard 12-lead ECG uses 10 electrodes: four on the limbs (just above the wrists and ankles) and six on the chest. V1 and V2 go in the 4th intercostal space at the right and left sternal edges, V4 goes in the 5th intercostal space in the mid-clavicular line, V3 sits midway between V2 and V4, and V5 and V6 sit at the same horizontal level as V4 in the anterior and mid-axillary lines. In the UK, leads follow the IEC colour code: right arm red, left arm yellow, left leg green, right leg black.
Small errors matter. The Society for Cardiological Science and Technology (SCST) notes that V1 and V2 are frequently placed too high and V4 to V6 too low, and either error can produce waveform changes that look like real pathology. This guide sets out every electrode position as described in the SCST's Clinical Guidelines by Consensus: Recording a Standard 12-lead Electrocardiogram (September 2024), with colour codes for both the IEC and AHA conventions, a step-by-step recording method, the most common mistakes, troubleshooting for poor tracings and guidance on additional leads.
In this article
- What is a 12-lead ECG and why does electrode placement matter?
- How do you prepare a patient for a 12-lead ECG?
- Where do the four limb electrodes go?
- Where do the chest leads V1 to V6 go?
- What are the IEC and AHA ECG colour codes?
- How do you record a 12-lead ECG step by step?
- What are the most common ECG lead placement mistakes?
- How do you fix a noisy or poor-quality ECG?
- When do you need right-sided or posterior leads?
- Which part of the heart does each lead view?
- What ECG electrodes and machines should you use?
- Frequently asked questions
What is a 12-lead ECG and why does electrode placement matter?
A 12-lead electrocardiogram records the heart's electrical activity from 12 different viewpoints. Despite the name, it uses only 10 physical electrodes. The machine combines the signals from those electrodes to produce 12 leads, which fall into two groups:
| Lead group | Leads | Derived from | Plane viewed |
|---|---|---|---|
| Standard (bipolar) limb leads | I, II, III | RA, LA and LL electrodes | Frontal (vertical) |
| Augmented limb leads | aVR, aVL, aVF | RA, LA and LL electrodes | Frontal (vertical) |
| Precordial (chest) leads | V1 to V6 | Six chest electrodes | Horizontal (transverse) |
| Neutral electrode | None | RL electrode | Reference only; reduces interference |
Because each lead is a fixed viewpoint, moving an electrode changes the view. A chest electrode one intercostal space out of position can alter R-wave progression, create an rSr' pattern in V1 that resembles right bundle branch block, or change ST segments and T waves enough to suggest ischaemia that is not there. Serial ECGs are only comparable if electrodes go in the same anatomical positions every time. Accurate placement against fixed bony landmarks is the single most controllable factor in ECG quality.
How do you prepare a patient for a 12-lead ECG?
The SCST guideline sets out preparation in a specific order. Good preparation reduces artefact and avoids repeat recordings.
Check identity. Confirm at least two unique identifiers, such as full name and date of birth, and enter them on the machine. The SCST warns that some machines keep the previous patient's details if new ones are not entered, so always check the printout.
Explain and gain consent. Introduce yourself, explain what the test involves, how much clothing needs to be removed and that adhesive electrodes will be used. Verbal consent is acceptable. A formal chaperone should be offered; the SCST does not recommend relatives or carers as chaperones, and if a chaperone is declined the offer and refusal should be recorded.
Position the patient. The SCST recommends a semi-recumbent position of about 45 degrees, with the limbs supported on the couch to reduce muscle tension. It notes there is no evidence that inclinations between flat and 45 degrees significantly change the ECG, but any significant variation from the standard position should be documented on the recording.
Prepare the skin. Clean each site with mild soap and water or an alcohol wipe and let it dry. Lightly exfoliate if needed using a paper towel, gauze swab or abrasive tape. Remove chest hair where necessary, with verbal consent, using a battery or single-use razor, and dispose of the blade in a sharps bin. Take care with fragile or broken skin.
Check your consumables. Electrodes with dried gel give high impedance and a poor trace. Use electrodes within their expiry date, keep opened pouches sealed and use a type designed for resting ECGs (tab or resting electrodes) rather than long-wear monitoring electrodes.

Adult resting electrodes designed for routine 12-lead recordings, in a bulk pack of 500 for high-throughput clinics.
Where do the four limb electrodes go?
The SCST guideline places limb electrodes on the limbs, proximal to the wrists and ankles, wherever possible. They produce the six frontal-plane leads (I, II, III, aVR, aVL and aVF). Place each electrode on a flat, fleshy area rather than directly over bone.
| Electrode | IEC label and colour (UK) | AHA label and colour (US) | Position (SCST) |
|---|---|---|---|
| Right arm | R, red | RA, white | Right forearm, proximal to the wrist |
| Left arm | L, yellow | LA, black | Left forearm, proximal to the wrist |
| Left leg | F, green | LL, red | Left lower leg, proximal to the ankle |
| Right leg (neutral) | N, black | RL, green | Right lower leg, proximal to the ankle |
Do not move limb electrodes onto the torso for a standard resting ECG. The SCST states that torso placement significantly alters wave amplitudes. Moving electrodes up the limbs should be avoided unless the patient has a significant tremor or an amputation. If you do have to move them, move the matching electrode on the other side to the same level, and annotate the ECG with the non-standard position.
A torso-based arrangement (the Mason-Likar modification, with arm electrodes below the clavicles and leg electrodes on the lower abdomen) is used during exercise stress testing and some continuous monitoring to reduce movement artefact. Recordings made this way are not directly comparable with a standard resting 12-lead ECG and must be clearly labelled.
Where do the chest leads V1 to V6 go?
The six precordial electrodes need the most care because they are placed against specific rib spaces. These are the SCST positions:
| Lead | Position | Placement tip |
|---|---|---|
| V1 | 4th intercostal space, right sternal edge | Count down from the angle of Louis; the most frequent error is placing it too high |
| V2 | 4th intercostal space, left sternal edge | Count the left side separately; rib spaces can be offset |
| V3 | Midway between V2 and V4 | Place V4 first, then position V3 |
| V4 | 5th intercostal space, mid-clavicular line | The reference level for V5 and V6 |
| V5 | Left anterior axillary line, same horizontal level as V4 | Do not follow the 5th intercostal space round the chest |
| V6 | Left mid-axillary line, same horizontal level as V4 and V5 | Should sit level with V4, not lower |
How to find the 4th intercostal space
The SCST method uses the manubriosternal joint, also called the sternal angle or angle of Louis, as the reference point:
- Run a finger down the sternum from the suprasternal notch until you feel a horizontal bony ridge. This is the angle of Louis, where the 2nd rib joins the sternum.
- Slide your finger down and to the patient's right to find the 2nd intercostal space, just below the 2nd rib.
- Count down to the 3rd and then the 4th intercostal space. Place V1 at the right sternal edge.
- Repeat the count on the left side for V2. Do not assume the spaces line up across the sternum.
- Continue down to the 5th intercostal space and follow it out to the mid-clavicular line for V4.
- Place V3 midway between V2 and V4, then V5 and V6 in horizontal alignment with V4.
Do not count the small gap between the clavicle and the 1st rib as the 1st intercostal space. That mistake moves every chest electrode one space too high.
Placing chest electrodes in women
When breast tissue overlies the correct positions, the SCST convention is to place V4, V5 and V6 beneath the left breast, on the chest wall. Lift the breast with care and sensitivity, for example using the back of the hand, and explain what you are doing. The guideline acknowledges that some evidence suggests placing electrodes on the breast may not significantly attenuate the signal, but recommends keeping to the convention until there is stronger evidence. It also advises extra care when interpreting ECGs from women with breast implants.
What are the IEC and AHA ECG colour codes?
Two colour conventions are in use. The International Electrotechnical Commission (IEC) system is used in the UK and most of Europe; the American Heart Association (AHA) system is used mainly in North America. Electrode positions are identical in both; only the labels and colours differ. If you buy imported cables or machines, check which convention they use before your first recording. The SCST also notes that colours can vary between manufacturers, so always read the label on the lead.
| Position | IEC label | IEC colour | AHA label | AHA colour |
|---|---|---|---|---|
| Right arm | R | Red | RA | White |
| Left arm | L | Yellow | LA | Black |
| Left leg | F | Green | LL | Red |
| Right leg | N | Black | RL | Green |
| V1 | C1 | Red | V1 | Red |
| V2 | C2 | Yellow | V2 | Yellow |
| V3 | C3 | Green | V3 | Green |
| V4 | C4 | Brown | V4 | Blue |
| V5 | C5 | Black | V5 | Orange |
| V6 | C6 | Violet | V6 | Purple |
Memory aids. For IEC limb leads, start at the right arm and go clockwise round the patient (right arm, left arm, left leg, right leg): "Ride Your Green Bike" gives red, yellow, green, black. For IEC chest leads, the sequence red, yellow, green, brown, black, violet is the same order as the limb colours with brown and violet added. For the AHA system, "white on the right, smoke over fire" places white on the right arm and black (smoke) on the left arm above red (fire) on the left leg.
How do you record a 12-lead ECG step by step?
Once the patient is prepared, work through the recording in the same order every time. Consistency is what makes serial ECGs comparable.
- Confirm identity and consent, enter patient details on the machine and offer a chaperone.
- Position the patient semi-recumbent at about 45 degrees with arms and legs supported.
- Prepare the skin at each site: clean, dry, lightly abrade and remove hair if needed.
- Attach the four limb electrodes just above the wrists and ankles, on the inner aspect of the limbs.
- Locate the angle of Louis and count down to the 4th intercostal space. Place V1 (right sternal edge) and V2 (left sternal edge).
- Place V4 in the 5th intercostal space at the mid-clavicular line, then V3 midway between V2 and V4.
- Place V5 and V6 level with V4 at the anterior and mid-axillary lines.
- Connect the lead wires, checking each label and colour against its electrode. Make sure clips are fully attached and cables are not pulling on the electrodes.
- Ask the patient to relax, breathe normally and stay still and quiet.
- Record with standard settings: 25 mm/s paper speed and 10 mm/mV gain, with a rhythm strip. The SCST recommends recording first with the muscle filter off.
- Check the tracing before removing the electrodes. If complexes overlap, gain can be reduced to 5 mm/mV; any change must be marked on the ECG.
- Annotate any non-standard electrode or patient position, filter use, gain change and symptoms at the time of recording (for example chest pain or palpitations).
The SCST's standard settings are summarised below:
| Setting | Standard | When to change |
|---|---|---|
| Paper speed | 25 mm/s | Any change must be marked on the ECG |
| Gain | 10 mm/mV | 5 mm/mV if complexes overlap; mark the change |
| Bandwidth (auto mode) | 0.67 to 150 Hz, filter off | Repeat at 0.67 to 40 Hz only if muscle interference is present; label the filter |
| Bandwidth (manual mode) | 0.05 to 150 Hz | Not applicable |
| Mains (50 Hz) filter | Off | Find and remove the source of interference first |
| Children | High-frequency cut-off no lower than 250 Hz | Adults: no lower than 150 Hz |
Filters reduce interference but also distort the ECG, which is why the SCST asks for the unfiltered recording first and clear labelling of any filtered repeat.
What are the most common ECG lead placement mistakes?
1. V1 and V2 placed too high. The SCST identifies this as a frequent error. Recording V1 and V2 in the 2nd or 3rd intercostal space can produce an rSr' pattern in V1 that mimics right bundle branch block, T-wave inversion or apparent ST changes in the anterior leads. Always count down from the angle of Louis rather than estimating from the nipple line or the top of the sternum.
2. V4 to V6 placed too low. The SCST also notes that V4 to V6 are frequently placed too low. A common cause is following the 5th intercostal space as it curves downwards round the chest wall, rather than keeping V5 and V6 level with V4. This alters R-wave progression and lateral lead amplitudes.
3. V3 placed by eye before V4. V3 is defined as midway between V2 and V4, so it can only be placed accurately once V4 is in position.
4. Arm lead reversal. Swapping the right and left arm cables inverts lead I (negative P wave and QRS complex), swaps leads II and III and swaps aVR with aVL. The trace can look plausible, and the pattern resembles dextrocardia. If lead I is unexpectedly negative, check the cables before drawing any conclusion. Trace each wire from the electrode back to its label.
5. Limb electrodes on the torso. Using torso positions for a standard resting ECG changes the frontal axis and wave amplitudes. Keep electrodes on the limbs unless there is a documented reason.
6. Poor skin contact or old electrodes. Oily skin, hair, lotion and dried electrode gel all increase impedance, causing baseline wander, noise or a flat trace on one channel.
7. Missing annotation. Any variation (patient sitting upright, limb electrodes moved for tremor, filter on, gain halved) changes how the ECG should be read. If it is not on the recording, the person reporting it cannot allow for it.
How do you fix a noisy or poor-quality ECG?
Identify the type of artefact first, then work through the likely causes.
| Problem | What it looks like | Likely causes | What to do |
|---|---|---|---|
| Baseline wander | Slow drift of the baseline up and down | Poor electrode contact, breathing movement, cables pulling | Re-prepare the skin, replace electrodes, support cables, ask the patient to breathe gently |
| Mains interference | Regular fine fuzz at 50 Hz | Nearby equipment or power cables, poor neutral electrode contact | Move power leads away, unplug unnecessary equipment, check the right leg electrode, keep the patient off metal couch parts |
| Muscle (somatic) artefact | Irregular, jittery baseline | Cold, anxious or tense patient, unsupported limbs, tremor | Warm and reassure the patient, support the limbs; if it persists, repeat with the 40 Hz filter and label it |
| Flat or absent lead | One channel shows no signal | Disconnected clip, broken lead wire, dry electrode | Check the clip, replace the electrode, test or replace the lead |
| Unexpected negative lead I | Inverted P wave and QRS in lead I | Arm leads reversed, or true dextrocardia | Check cable positions; if correct, consider dextrocardia (see below) |
For patients with a significant tremor, the SCST allows limb electrodes to be moved up the limbs. Move both sides equally and annotate the recording.

Compact 34mm x 23mm tab electrodes with a gentle solid gel for resting ECGs, supplied ten per liner.
When do you need right-sided or posterior leads?
Additional leads are recorded on clinical request or according to local policy. Always annotate them clearly, because the machine will usually still label the channels V1 to V6.
Right-sided leads (V3R to V6R). These mirror the standard V3 to V6 positions on the right side of the chest, using the same landmarks. V4R, for example, is in the 5th intercostal space in the right mid-clavicular line. Right-sided leads are commonly requested when an inferior myocardial infarction is suspected, to look for right ventricular involvement, which affects management.
Dextrocardia. The SCST advises suspecting dextrocardia when lead I shows negative P waves and QRS complexes with no technical error, especially with poor R-wave progression across V1 to V6. Record a second ECG with V3R to V6R on the right side, keeping V1 and V2 in place, and annotate it clearly. The arm connections can also be swapped for this recording, again with very clear annotation.
Posterior leads (V7 to V9). All three are placed at the same horizontal level as V4:
- V7: left posterior axillary line
- V8: left mid-scapular line
- V9: left spinal border
If the machine cannot record 15 leads, the SCST suggests recording a standard 12-lead ECG first, then moving the V4 to V6 electrodes to the V7 to V9 positions for a second, annotated recording.
Children. The method is the same as for adults, with a gentle approach and smaller electrodes where needed. Semi-recumbent is preferred, but sitting is acceptable if it avoids distress. V4R, V5R and V7 may be recorded on request; the SCST notes that V4R is commonly used and V3 often omitted in infants up to one year old.
Which part of the heart does each lead view?
Knowing which leads look at which area helps you recognise when a pattern is anatomically consistent, and when it might be a placement artefact. Changes confined to anatomically contiguous leads point towards a real regional problem; changes in a single misplaced lead often do not. Coronary anatomy varies between individuals, so the artery column shows the usual supply only.
| Region | Leads | Usual coronary supply |
|---|---|---|
| Inferior | II, III, aVF | Right coronary artery (most people) or circumflex |
| Septal | V1, V2 | Left anterior descending (septal branches) |
| Anterior | V3, V4 | Left anterior descending |
| Lateral | I, aVL, V5, V6 | Circumflex or diagonal branches of the LAD |
| Right ventricle | V4R (additional lead), V1 | Right coronary artery |
| Posterior | V7 to V9 (additional leads); reciprocal changes in V1 to V3 | Right coronary or circumflex |
Interpretation should always be done by a clinician trained to report ECGs, in line with local policy and the SCST's separate reporting standards.
What ECG electrodes and machines should you use?
Electrodes. For resting 12-lead ECGs, most UK clinics use either tab electrodes (small, flat electrodes connected with crocodile or alligator clips) or snap/stud electrodes, depending on the machine's cables. Tab electrodes such as the Welch Allyn resting tab electrodes (Ag/AgCl, pack of 500, £45.49) and the Ambu Blue Sensor Tab electrodes (£4.79 per 100) are quick to apply and economical for routine recording. For tab electrodes you need clips such as the Seca universal alligator clips (pack of 10, £26.69). General-purpose disposable electrodes such as Skintact disposable ECG electrodes (x 500, £18.79) suit resting and monitoring use. For Holter and other 24-hour recording, use long-wear electrodes such as the Skintact F-TB-3 24hr electrodes (pack of 3, £1.99) rather than resting tabs. The full range is in our ECG collection.
Machines. When choosing a 12-lead ECG machine for a GP surgery or clinic, look at integration with your clinical system (EMIS, SystmOne, Vision), interpretation software, screen size, weight and paper format. Some machines also show an on-screen electrode placement guide and lead-reversal alerts, which can help less experienced staff.
| Feature | seca CardioPad-2 | seca CT8000i-2 |
|---|---|---|
| Display | 8" colour multi-touch screen with anatomical electrode placement guide | 5" colour LCD |
| Weight | 1.1kg including paper | 1.9kg including paper |
| Data transfer | Automatic PDF transfer via Wi-Fi or LAN | Automatic PDF export to USB |
| Placement checks | Lead inversion and incorrect placement alerts before capture | On-screen signal check and ECG preview |
| Internal memory | Up to 350 ECGs | Up to 100 ECGs |
| QTc formulas | Bazett, Fridericia, Framingham, Hodges | Bazett, Fridericia, Framingham, Hodges |
| Price | From £2,809.59 | From £1,603.33 |
Other resting ECG machines, including the seca CT8000p-2, are listed in the ECG collection, along with compatible thermal paper.

1.1kg touch-screen ECG with an on-screen electrode placement guide, lead-inversion alerts and Wi-Fi transfer to EMIS, SystmOne and Vision.

Compact 1.9kg ECG with a 5" colour screen, built-in thermal printer, ETM interpretation and QTc measurement.
ECG electrodes, machines and paper
Resting and monitoring electrodes, 12-lead ECG machines and thermal paper for GP surgeries and clinics. Free UK delivery on orders over £35.
Shop ECG Supplies →Frequently asked questions
How many electrodes does a 12-lead ECG use?
A 12-lead ECG uses 10 electrodes: four limb electrodes (right arm, left arm, left leg and right leg) and six chest electrodes (V1 to V6). The machine combines their signals to produce 12 leads, or views, of the heart.
Where does V1 go on an ECG?
V1 goes in the 4th intercostal space at the right sternal edge. Find it by locating the angle of Louis (where the 2nd rib meets the sternum), moving into the 2nd intercostal space and counting down to the 4th. V2 sits in the same space at the left sternal edge.
Should V5 and V6 follow the 5th intercostal space?
No. V5 and V6 are placed at the same horizontal level as V4, in the anterior and mid-axillary lines respectively. Following the intercostal space round the chest places them too low, which the SCST lists as a common error.
What is the difference between IEC and AHA ECG colour codes?
The positions are the same; only the colours and labels differ. Under the IEC system used in the UK, the limb leads are red (right arm), yellow (left arm), green (left leg) and black (right leg), and the chest leads run red, yellow, green, brown, black, violet. The AHA system used in North America has white, black, red and green limb leads and red, yellow, green, blue, orange, purple chest leads.
Where do ECG electrodes go on women?
The landmarks are the same. Where breast tissue overlies the correct positions, the SCST convention is to place V4, V5 and V6 beneath the left breast on the chest wall, lifting the breast carefully, for example with the back of the hand. Offer a chaperone and explain each step.
Can limb electrodes be placed on the chest or shoulders?
Not for a standard resting 12-lead ECG. The SCST states that limb electrodes must not be placed on the torso because it significantly alters wave amplitudes. They should only be moved up the limbs for significant tremor or amputation, and the change must be annotated. Torso placement is used for exercise testing, where it is labelled as such.
What position should the patient be in for an ECG?
The SCST recommends a semi-recumbent position of about 45 degrees with the arms and legs supported. Any significant variation, such as recording a patient sitting upright in a chair, should be noted on the ECG.
What paper speed and gain should a standard ECG use?
The standard is 25 mm/s and 10 mm/mV, with a rhythm strip. Gain can be reduced to 5 mm/mV if complexes overlap, and any change to speed, gain or filters must be marked on the ECG.
Disclaimer: This guide is for general information for trained healthcare staff and summarises the SCST Clinical Guidelines by Consensus: Recording a Standard 12-lead Electrocardiogram (September 2024). It is not a substitute for formal ECG training, your organisation's policies or the current SCST guideline. ECGs should be interpreted by an appropriately trained clinician. Always check the latest guidance and the manufacturer's instructions for your machine and cables, and speak to your clinical lead if you are unsure.